Abstract

"Stress Taper" Strategy in Plate Fixation Increases Failure Strength in a Cadaveric Femur Model

Purpose:To potentially limit peri-implant fractures, our institution commonly implements a fracture fixation strategy known as a “stress taper,” in which the screw lengths toward the end of a construct are incrementally decreased. The premise is to dissipate the stress as one moves proximally in order to avoid a focal stress-riser when loaded. The aim of this study was to determine if the stress taper strategy increases torsional strength and fails in a simpler fracture pattern than the bicortical locking construct when biomechanically tested in a cadaveric femur model. Methods: Including an assessment of the failure mechanism Seven matched pairs of cadaveric femora were randomly assigned to 1 of 2 distal femur fixation groups: plating with stress taper strategy or bicortical fixation. Both strategies used locking plates and the same number of locking screws in the same positions. Specimens were first cyclically loaded in axial compression and torsion. Specimens were then axially rotated to failure under 800 N of compression. Peak torque at failure and degrees of rotation at failure were calculated and compared using paired t tests. Fractures were categorized with the assistance of fluoroscopy according to the OTA 32.


Author(s): Jeremy Raducha

Abstract | Full-Text | PDF

Share This Article